原稿を提出する arrow_forward arrow_forward ..

Long-Term Monitoring of Intracardiac Pressure at Home using a Wireless Implantable Sensor

Abstract

Henrik Casimir Ahn, MD, PhD and Jacek Baranowski, MD, PhD

Background: We have previously reported results from an initial study on a new implantable intracardiac wireless hemodynamic monitor in man, showing favourable results regarding accuracy of recordings over time and safety profile. The aim of this study was to follow up the intervention group of 16 patients with left atrial implants over a longer period to determine the systems longevity and dependability.
Material: The Titan™ monitoring system using a miniature pressure sensor was implanted in 16 patients undergoing open chest cardiac surgery. After discharge the system was used for left atrial pressure monitoring at home.
Findings: No major or minor adverse events were observed. As of July 2018, the average age of functioning left atrial implants providing high-quality hemodynamic curves was 3.75 years, ranging from 3.25 years to 5 years.
Conclusion: The system was reliable, provided high-quality pressure curves and was easy to use. Recalibration using the non-invasive Valsalva maneuver was accurate and easy to perform.

免責事項: この要約は人工知能ツールを使用して翻訳されており、まだレビューまたは確認されていません